Name/Model: Universal Testing Machine Instron 5566A

Manufacturer: Instron, Norwood, MA

Specifications: Electromechanical load frames are designed to apply a load to a test specimen via the moving crosshead.

Software version: Bluehill 3.0

Applications: Tension, Compression, Tension Creep Relaxation, Compression Creep Relaxation, Tension Profile, Compression Profile, Peel/Tear/Friction, Flexure, Metals (Yield), Flexure Creep Relaxation

Generic SOP for Universal Testing Machine Instron 5566A

Sample requirements:

  • Prepare your samples according to the test requirements.
  • Different test has diffrent requirements on sample demension and fixture.

Preparation:

  • Check the power switch of instron and printer
  • Check the status of computer, log in as administrator.
  • Check the load cell cable is connected and in the proper load range as you need.
  • Prepare fixture for your specimens.

Testing:

  • Run BlueHill 3 from computer.
  • Load your testing method and choose testing filename, save to your own folder. Each group pf specimens can be included in one data file.
  • Input speciment information if needed.
  • Mount your specimens in the proper fixture.
  • Align the load cell with your specimens. Secure the fixture.
  • Set the extension reset before the test. Set the frame safety sensors for upper and lower limit.
  • For testing of load to failure, install proper protection around the fixture and wear safety googoles for yourself.
  • Click “start” to test.
  • Collect finished specimens and save for further fractograph analysis.
  • Save the results after testing one group specimens. Do not mix different groups of specimens in one data file.

Change load cells:

  • Unscrew the three bolts with Allen wrench.
  • Unplug the cable conector.
  • Find the empty case with the same labelled range as current load cell from cabinet. Leave it open on bentch.
  • Carefully remove the load cell from crosshead and put into the case. Lock the case and put it back to cabinet.
  • Find the case with load cell you need. Open the case and take out of the load cell.
  • Mount the load cell on crooshead with bolts. Connect the cable
  • Warning: You must return load cell back to case when using 10 N or 100 N load cell.

Finish:

  • Switch off the power of Instron. You may leave it on and we do not see any harm by doing so.
  • Clean the debris.
  • Sign off logbook.